Chrome Extension
WeChat Mini Program
Use on ChatGLM

Magnetic resonance imaging safety in patients with abandoned or functioning epicardial pacing leads

European Radiology(2022)

Cited 10|Views12
No score
Abstract
Objectives The European Society of Cardiology Guidelines on cardiac pacing from 2021 allow magnetic resonance imaging (MRI) in patients with cardiac implantable electronic devices (CIEDs) but do not recommend MRI in patients with epicardial pacing leads. The clinical dilemma remains whether performing an MRI in patients with CIED and epicardial leads is safe. We aimed to evaluate the safety of performing an MRI in patients with CIED and abandoned or functioning epicardial pacing leads. Methods We included all adult patients who underwent clinically indicated MRIs with CIED and functioning or abandoned epicardial leads in a single tertiary hospital between November 2011 and October 2019. The data were retrospectively collected. Results Twenty-six MRIs were performed on 17 patients with functioning or abandoned epicardial pacing leads. Sixty-nine percent of the MRI scans (18/26) were conducted on patients with functioning epicardial pacing leads. A definite adverse event occurred in one MRI scan. This was a transient elevation of the pacing threshold in a patient with a functioning epicardial ventricular pacing lead implanted 29 years previously. An irreversible atrial pacing lead impedance elevation was detected 6 months after the MRI in another patient; the association with the previous MRI remained unclear. No adverse events were detected in MRIs performed on patients with modern (implanted in 2000 or later) functioning epicardial leads. Conclusions MRIs in patients with CIED and modern functioning epicardial pacing leads were performed without detectable adverse events. Further large-scale studies are necessary to confirm MRI safety in patients with epicardial pacing leads. Key Points • Currently, MRI in patients with cardiac implantable electronic devices (CIEDs) and functioning or abandoned epicardial pacing leads is not recommended. • MRIs in patients with CIED and modern functioning epicardial leads (implanted in 2000 or later) were performed without detectable adverse events in our patient cohort. • Allowing MRI in patients with epicardial pacing leads may significantly improve the diagnostic work-up, especially in specific patient groups, such as patients with congenital heart disease.
More
Translated text
Key words
Magnetic resonance imaging, Pacemaker, artificial, Arrhythmias cardiac, Electrodes, implanted
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined