Determination of Phenolic Compounds in Beverages by Three-Flow Channel Isocratic HPLC with Electrochemical Detections Using a Column-Switching Technique

CHEMICAL & PHARMACEUTICAL BULLETIN(2022)

引用 4|浏览2
暂无评分
摘要
For the quantitative analysis of phenolic compounds in beverage samples, a three-flow channel isocratic HPLC with electrochemical detection (3LC-ECD) system was devised using a column-switching technique. Phenolic compounds with significantly different hydrophobicity (the range of calculated logP: -0.77 to 3.02) were simultaneously measured to draw three chromatograms by the 3LC-ECD; the peaks of gallic acid (GA), protocatechuic acid (PCA), and gallocatechin (GC) were observed at electrochemical detector 1 (D1) within 42min, the peaks of procyanidin B3 (B3), epigallocatechin (EGC), catechin (C), epicatechin (EC), procyanidin B2 (B2), ethyl gallate (Eg), and epigallocatechin gallate (EGCg) were observed at D2 within 50min, and the peaks of epicatechin gallate (Erg), gallocatechin gallate (Grg), catechin gallate (Cg), propyl gallate (Pg), and resveratrol (RVT) were observed at D3 within 70min. The relationships between the phenolic compound concentrations and their chromatographic peak heights gave good linearity with correlation coefficients of more than 0.998. The detection limits of the phenolic compounds by the 3LC-ECD ranged from 0.6 to 3.0 mu g/L. Further, the phenolic compound concentrations of commercially available teas and wines were determined with a relative standard deviation (RSD) of less than 4.9% (n= 6), and their recoveries ranged from 91 to 109%. These results indicate that the 3LC-ECD system provided an accurate, precise, and specific determination of the phenolic compounds in beverages without affecting the matrices derived from these samples.
更多
查看译文
关键词
electrochemical detection, column switching, phenolic compound, tea, wine
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要