High-quality quasi-parallel X-ray beam obtained by a parabolic monocapillary X-ray lens with a square beam stop

JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY(2022)

引用 1|浏览13
暂无评分
摘要
Parabolic monocapillary X-ray lens (PMXRL) is an ideal optical device for constraining the point divergent Xray beams to quasi-parallel beams, but the overlap of direct X-rays and reflected X-rays through PMXRL deteriorates the outgoing beam divergence. Aiming to solve this problem, this study designs and tests a square-shaped lead occluder (SSLO) embedded in PMXRL to block the direct X-rays passing through the PMXRL. Python simulations are employed to determine the geometric parameters of the SSLO as well as the optimal position of the SSLO in the PMXRL according to our proposed model. The PMXRL with a conic parameter p of 0.000939mm and a length L of 60.8mm is manufactured and the SSLO with a size of 0.472mmx0.472mmx3.4mm is embedded into it. An optical path system based on this PMXRL is built to measure the divergence of the outgoing X-ray beam. The experimental results show that the quasi-parallel X-ray beam reaches a divergence of 0.36 mrad in the range from 15-45mm at the PMXRL outlet. This divergence is 10 times lower than the theoretical divergence without SSLO. Our work provides an alternative method for obtaining highly parallel X-ray beam and is beneficial to generate or facilitate new applications of monocapillary optics in X-ray technology.
更多
查看译文
关键词
Monocapillary optics, quasi-parallel beam, beam stop, beam divergence