Chrome Extension
WeChat Mini Program
Use on ChatGLM

Comparison of MLC positioning deviations using log files and establishment of specific assessment parameters for different accelerators with IMRT and VMAT

Radiation Oncology(2022)

Cited 1|Views3
No score
Abstract
Background and purpose The study evaluated the differences in leaf positioning deviations by the log files of three advanced accelerators with two delivery techniques, and established specific assessment parameters of leaf positioning deviations for different types of accelerators. Methods A total of 420 treatment plans with 5 consecutive treatment log files were collected from the Trilogy, TrueBeam and Halcyon accelerators. Millennium MLC was equipped on the Trilogy and TrueBeam accelerators. A jawless design and dual-layer MLC were adopted on the Halcyon accelerator. 70 IMRT and 70 VMAT plans were selected randomly on each accelerator. The treatment sites of all plans included head and neck, chest, breast, pelvis and other sites. The parsing tasks for 2100 log files were proceeded by SunCheck software from Sun Nuclear Corporation. The maximum leaf root mean square (RMS) errors, 95th percentile errors and percentages of different leaf positioning errors were statistically analyzed. The correlations between these evaluation parameters and accelerator performance parameters (maximum leaf speed, mean leaf speed, gantry and arc angle) were analyzed. Results The average maximum leaf RMS errors of the Trilogy in the IMRT and VMAT plans were 0.44 ± 0.09 mm and 0.79 ± 0.07 mm, respectively, which were higher than the TrueBeam's 0.03 ± 0.01 mm, 0.03 ± 0.01 mm and the Halcyon's 0.05 ± 0.01 mm, 0.07 ± 0.01 mm. Similar data results were shown in the 95th percentile error. The maximum leaf RMS errors were strongly correlated with the 95th percentile errors (Pearson index > 0.5). The leaf positioning deviations in VMAT were higher than those in IMRT for all accelerators. In TrueBeam and Halcyon, leaf position errors above 1 mm were not found in IMRT and VMAT plans. The main influencing factor of leaf positioning deviation was the leaf speed, which has no strong correlation with gantry and arc angles. Conclusions Compared with the quality assurance guidelines, the MLC positioning deviations tolerances of the three accelerators should be tightened. For both IMRT and VMAT techniques, the 95th percentile error and the maximum RMS error are suggested to be tightened to 1.5 and 1 mm respectively for the Trilogy accelerator. In TrueBeam and Halcyon accelerators, the 95th percentile error and maximum RMS error of 1 and 0.5 mm, respectively, are considered appropriate.
More
Translated text
Key words
MLC positioning error, Trajectory log file, Dynalog, IMRT, VMAT, Trilogy, TrueBeam, Halcyon
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined