An electrochemical sensor based on a Co3O4-ERGO nanocomposite modified screen-printed electrode for detection of uric acid in artificial saliva

ANALYTICAL METHODS(2021)

引用 22|浏览4
暂无评分
摘要
In this study, we report the fabrication of a nanocomposite consisting of Co3O4 nanoparticles (Co3O4 NPs) and electrochemically reduced graphene oxide (ERGO) on a screen-printed electrode (SPE) and its sensing performance in the electrochemical detection of uric acid (UA). The surface modification of the electrode was confirmed by using a variety of characterization techniques (FE-SEM, XRD, AFM, EDX, WCA, FTIR, and Raman spectroscopy). In addition, the surface modification was electrochemically characterized step by step through CV, EIS and DPV techniques, and the results showed that the Co3O4-ERGO nanocomposite exhibited highly sensitive and selective sensing performance towards the oxidation of UA in 0.1 M (pH 7.0) phosphate buffer solution (PBS). The sensor (Co3O4-ERGO/SPE) signals were observed to be linear to the UA concentration in the range of 5 mu M to 500 mu M (R-2 = 0.9985). After revealing its other performance characteristics, such as repeatability, reproducibility, stability, sensitivity, and selectivity, the sensor was successfully applied to the analysis of UA in artificial saliva samples.
更多
查看译文
关键词
electrochemical sensor,electrode,artificial saliva,uric acid,screen-printed