Right Heart Morphology of Candidate Patients for Transcatheter Tricuspid Valve Interventions

Cardiovascular Engineering and Technology(2021)

Cited 3|Views11
No score
Abstract
Purpose This study quantitatively evaluated the phasic right heart morphology of candidate patients for a transcatheter tricuspid valve intervention (N=32) and of subjects with trace to no tricuspid regurgitation ( N = 14). Methods Cardiac computed tomography angiography (CCTA) and transthoracic/transesophageal echocardiography (TTE/TEE) images were analyzed using dedicated research and clinical software. Using CCTA, the phasic right atrial and ventricular volumes, annulus dimensions, annulus-to-right coronary artery (RCA) distances, circumferential topography of the annular tissue shelf, vena cava dimensions (inferior and superior), vena cava positions, axis angles, and annular excursions were quantified. Using TTE/TEE, leaflet geometry, regurgitation, hemodynamics, and heart function were quantified. Measurements within and between groups were quantitatively compared with regression analyses to explore relationships between right heart features. Results The phasic position and orientation of the vena cava and the circumferential topography of the annular tissue shelf were quantitatively presented for the first time. The candidate patient group exhibited greater chamber dimensions, enlarged vena cava, distended vena cava positions, positional shallowing of the annular tissue shelf, geometric annular distortion, leaflet distention, moderate or greater regurgitation, and impaired ventricular function. Atrial volume correlated strongly with directional vena cava positions as well as with annular dimensions. Annulus-to-RCA distances and annular excursions were comparable between groups. Conclusions This study provides new and further insight to the right heart morphology and functional characteristics of candidate patients for a transcatheter tricuspid valve intervention. These data provide a platform from which these patients can continue to be better understood for further improving transcatheter system design and use.
More
Translated text
Key words
Heart,Tricuspid valve,Tricuspid regurgitation,Anatomy,Morphology,Computed tomography,Medical device,Design,Transcatheter therapies,Interventional cardiology,Medical imaging,Modeling
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined