Validation and verification of the GeneFinder™ COVID-19 Plus RealAmp kit on the ELITe InGenius® instrument
Journal of Virological Methods(2022)
摘要
•The GeneFinder CoViD19 assay has no cross-reactivity with other Coronavirusses.•Limit of detection for the GeneFinder CoViD19 N-gene was 50 digital PCR copies/mL.•Clinical results were comparable between the GeneFinder CoViD19 and in-house qPCR.
更多查看译文
关键词
SARS-CoV-2,GeneFinder™ COVID-19 PLUS RealAmp kit,ELITe InGenius® instrument
AI 理解论文