High-Performance NMHC Detection Enabled by a Perylene Bisimide-Cored Metallacycle Complex-Based Fluorescent Film Sensor

ANALYTICAL CHEMISTRY(2021)

引用 6|浏览12
暂无评分
摘要
Non-methane hydrocarbons (NMHCs) can serve as precursors of ozone and photochemical smog, and hence their highly efficient detection is of great importance for air quality monitoring. Here, we synthesized a new fluorescent perylene bisimide (PBI)-cored metallacycle complex through coordination-driven self-assembly and used it for the production of a fluorescent film sensor. The unique rectangular structure of the developed fluorophore endows the sensor with enhanced sensing performance and discriminability to n-alkanes (C5-10). Specifically, the experimental detection limits for n-pentane, n-hexane, and n-decane are 39, 7, and 1.4 mg/m(3), respectively, and the corresponding linear ranges are from 39 to 2546, 7 to 1745, and 1.4 to 85 mg/m(3), respectively. Moreover, the sensing is fully reversible. In tandem with a gas chromatographic separation system, the film sensor showed comparable detection ability for the n- alkanes with a commercial flame ionization detector (FID), while the film sensor needs no hydrogen; it occupies a much smaller size (30 X 30 X 44 mm(3)) and consumes less energy (0.215 W). Further studies demonstrated that the developed sensor can be used for on-site and real-time quantification of NHMCs, laying the foundation for developing into a portable detector.
更多