Abstract
Angle calibrations are widely used in various fields of science and technology, while in the high-precision angle calibrations, a complete closure method which is complex and time-consuming is common. Therefore, in order to improve the measurement efficiency and maintain the accuracy of the complete closure method, an improved calibration method was proposed and verified by the calibration of a high-precision angle comparator with sub-arc-second level. Firstly, a basic principle and algorithm of angle calibration based on complete closure and symmetry connection theory was studied. Then, depending on the pre-established calibration system, the comparator was respectively calibrated by two calibration methods. Finally, by comparing E-n values of two calibration results, the effectiveness of the improved method was verified. The calibration results show that the angle comparator has a stable angle position error of 0.17 '' and a measurement uncertainty of 0.05 '' (k = 2). Through method comparisons, it was shown that the improved calibration method can greatly reduce calibration time and improve the calibration efficiency while ensuring the calibration accuracy, and with the decrease of measurement interval, the improvement of calibration efficiency was more obvious.