Development And Performance Verification Of A Tpc Polarimeter For High Energy X-Rays

SPACE TELESCOPES AND INSTRUMENTATION 2020: ULTRAVIOLET TO GAMMA RAY(2021)

Cited 0|Views2
No score
Abstract
We present the first application of a time projection chamber polarimeter to measure high energy X-ray polarization above 10 keV. The polarimeter is designed based on the PRAXyS soft X-ray polarimeter. The sealed gas is changed to a gas mixture of 60% argon and 40% dimethyl ether at 1 atm to be sensitive to high energy X-rays. The polarimeter performance is verified with linearly polarized, monochromatic X-rays at a synchrotron radiation facility, KEK Photon Factory BL-14A. The measured modulation factors are 42.4 +/- 0.6%, 50.4 +/- 0.6%, and 55.0 +/- 0.6% at 12, 14, and 16 keV, respectively, and the measured polarization angles are consistent with the expected values at all energies.
More
Translated text
Key words
X-ray astronomy, polarimetry, photoelectric polarimeter, Time Projection Chamber (TPC)