Evaluation of a sensitive, reasonable, and fast detection method for 55Fe in steel
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY(2021)
摘要
A pilot study to quantify 55Fe in steel from a reactor vessel of a nuclear power plant by accelerator mass spectrometry (AMS) without any chemical sample preparation was validated by liquid scintillation counting (LSC) and AMS after radiochemical separation. AMS reaches an uncertainty < 10% at the 1 kBq gFe−1 level within less than 10 min measuring time. The background was < 3 Bq gFe−1, presently limited by the short measurement time. The new instrumental AMS method for analysing 55Fe from neutron capture production is reasonable and fast compared to other analytical methods.
更多查看译文
关键词
Accelerator mass spectrometry,Nuclear decommissioning,Nuclear waste,55Fe,Radiochemical separation,Liquid scintillation counting