A New Leadless Pacemaker With Atrioventricular Synchronous Pacing Replacing A Still Working Vvi Leadless Pacemaker: A Case Report

EUROPEAN HEART JOURNAL-CASE REPORTS(2021)

Cited 1|Views5
No score
Abstract
Background A new intracardiac leadless pacemaker (ILP) has been developed to ensure atrioventricular (AV) synchrony (AVS) during ventricular pacing (VP). Recent studies have shown the feasibility and safety of accelerometer-based atrial sensing and an improvement in AVS among patients with atrioventricular block implanted with the Micra AV ILP (Medtronic Inc., Minneapolis, MN, USA). However, no data exists about the benefits of a VDD ILP in patients wearing a still working VVI Nanostim ILP (St Jude Medical, St Paul, MN, USA). We describe the feasibility of the procedure and the absence of device-related adverse events in the short-term follow-up.Case summary We present the case of a 72-year-old man implanted with a VVI ILP (Nanostim, St Jude Medical, St Paul, MN, USA) on May 2014, who has developed symptomatic high percentage of VVI asynchronous pacing and was treated with an upgrade to synchronous AV pacemaker (PM) ILP-Micra AV (Medtronic Inc., Minneapolis, MN, USA), which has improved symptoms and functional class.Discussion Intracardiac leadless pacemakers represent the best current option for patients requiring PM implantation who are at high risk of infection and bleeding. Our case shows that the new AVS ILP is a good alternative to VVI ILP in patients with sinus rhythm and a strong need for VP.
More
Translated text
Key words
Case report, Leadless pacemaker, Micra TPS, Atrioventricular block, Atrioventricular synchronous pacing, Pacemaker syndrome
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined