Calibration Setting Numbers For Dose Calibrators For The Pet Isotopes Mn-52, Cu-64, Br-76, Y-86, Zr-89, I-124

APPLIED RADIATION AND ISOTOPES(2016)

引用 9|浏览0
暂无评分
摘要
For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels. (C) 2016 Elsevier Ltd. All rights reserved.
更多
查看译文
关键词
Dose calibrator, Ionization chamber, Calibration setting number, Instrumentation, Radiation measurements, Mn-52, Cu-64, Br-76, Y-86, Zr-89, I-124