谷歌浏览器插件
订阅小程序
在清言上使用

Research on leak detection of vials based on tunable diode laser absorption spectroscopy technology

ADVANCED SENSOR SYSTEMS AND APPLICATIONS X(2020)

引用 1|浏览3
暂无评分
摘要
The off-line sampling destructive test is often used in the seal integrity test of vials, which has the disadvantages of long time consuming, low accuracy and high missing rate. In order to solve the above problems, the tunable diode laser absorption spectroscopy technology and wavelength modulation spectroscopy technology are used to measure the oxygen concentration in the vials. The ratio method of second harmonic and first harmonic is used to eliminate the influence of the rapid change of light intensity. The experimental results show that the system can complete the seal integrity detection of 300 vials per minute. This method belongs to the non-destructive online detection mode, which can realize the fast detection of vials without interference.
更多
查看译文
关键词
tunable diode laser absorption spectroscopy,wavelength modulation spectroscopy,vial,leak detection,oxygen
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要