Isobaric Identification Using Gas-Filled Time-Of-Flight Measurements In An Accelerator Mass Spectrometry

CHINESE PHYSICS LETTERS(2005)

Cited 0|Views2
No score
Abstract
A gas-filled time-of-flight (GF-TOF) detector has been built and developed to improve the ability of isobaric identification in accelerator mass spectrometry (AMS) measurements, and a time resolution (without gas filled) of better than 350ps is achieved. The GF-TOF detector is tested by means of measuring a standard AgCl (Cl-36/Cl = 7.6 x 10(-9)g/g) sample with the Cl-36 ion energy of 64, 49 and 33 MeV, respectively. Cl-36 and S-36 particles were successfully separated in the TOF spectra output from the GF-TOF detector. The comparison between the gas-filled time-of-flight method and the Delta E - E method is described. Some results relative to the GF-TOF method are given as well.
More
Translated text
Key words
mass spectrometry,gas-filled,time-of-flight