Beating-Heart Surgical Treatment Of Atrial Fibrillation With Microwave Ablation

Jg Maessen, Jfma Nijs, Jlrm Smeets,J Vainer,B Mochtar

ANNALS OF THORACIC SURGERY(2002)

Cited 55|Views1
No score
Abstract
Background. In this feasibility study, early results are presented of our first series of patients with microwave ablation for atrial fibrillation (AF) on the beating heart.Methods. From June 2001 until December 2001, a total of 24 patients underwent beating-heart epicardial ablation for AF. With a microwave antenna, the left and right pulmonary veins were isolated and connected to each other followed by amputation of the left atrial appendage. Subsequently, patients underwent either off-pump coronary artery bypass graft or valve surgery on pump. The mean age of the patients was 67.4 +/- 6 years. Three patients experienced paroxysmal atrial fibrillation and all others chronic AF. Mean left atrial diameter was 5.4 +/- 0.6 cm, and mean ablation time was 13 min.Results. All procedures but one were completed successfully on the beating heart. All patients were in sinus rhythm after the procedure. A total of 15 patients experienced periods with postoperative AF during hospital stay; 9 of these patients were discharged with AF. All patients received either sotalol or antiodarone. At latest follow-up (3 to 9 months), 20 of 23 patients were in sinus rhythm.Conclusions. With microwave ablation, electrical isolation of the pulmonary veins can be achieved epicardially without cardiopulmonary bypass support.
More
Translated text