Assessing the accuracy of a new hand hygiene monitoring device (SmartRub (R)): from the laboratory to clinical practice

ANTIMICROBIAL RESISTANCE AND INFECTION CONTROL(2021)

引用 0|浏览4
暂无评分
摘要
Background We developed SmartRub (R) powered by iQati (R), an electronic device composed of a wristband and an alcohol-based handrub pocket-sized dispenser that measures and provides feedback on the duration of hand friction and the volume poured during each hand hygiene action. We aimed to assess the accuracy of SmartRub (R). Methods The specificity, sensitivity, positive and negative predictive values (PPV and NPV) of SmartRub (R) were assessed in a 3-phased experiment: (1) laboratory-controlled conditions with volunteers; (2) pre-planned clinical path with volunteers and (3) real clinical conditions with healthcare workers. The accuracy of SmartRub (R) was evaluated by quantifying its ability to correctly capture true hand hygiene actions and to not record other actions performed while wearing the device. Results In the laboratory, 7 volunteers performed 816 actions. Overall sensitivity was 94.1% (95% CI 91.4-96.2%) with a PPV of 99.0% (95% CI 97.3-99.6%) and specificity was 99.0% (95% CI 97.5-99.7%) with a NPV of 94.4% (95% CI 91.9-96.1%). During the pre-planned clinical path phase, 13 volunteers performed 98 planned paths and a total of 967 actions were performed. Overall sensitivity was 94.6% (95% CI 92.2-96.5%) with a PPV of 84.3% (95% CI 81.6-86.7%) and specificity was 82.4% (95% CI 78.7-85.7%) with a NPV of 93.9% (95% CI 91.3-95.7%). During the real clinical conditions phase, 17 healthcare workers were observed for a total of 15 h and 3 min while they performed 485 actions. Sensitivity was 96.8% (95% CI 93.8-98.6%) with a PPV of 98.3% (95% CI 95.6-99.3%) and specificity was 98.3% (95% CI 95.7-99.5%) with a NPV of 96.8% (95% CI 93.9-98.4%). Conclusions Smartrub (R) is a highly reliable device for capturing hand hygiene actions under a range of conditions, from the laboratory to clinical care activities.
更多
查看译文
关键词
Hand hygiene, Hospital acquired infections, Electronic monitoring device, Sensitivity, Specificity
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要