An integrated system for automated measurement of airborne pollen based on electrostatic enrichment and image analysis with machine vision

Jia-Jing Yang, Christian Klinkenberg,Jian-Zhang Pan,Hans M Wyss,Jaap M J den Toonder,Qun Fang

TALANTA(2022)

引用 2|浏览12
暂无评分
摘要
Here we describe an automated and compact pollen detection system that integrates enrichment, in-situ detection and self-cleaning modules. The system can achieve continuous capture and enrichment of pollen grains in air samples by electrostatic adsorption. The captured pollen grains are imaged with a digital camera, and an automated image analysis based on machine vision is performed, which enables a quantification of the number of pollen particles as well as a preliminary classification into two types of pollen grains. In order to optimize and evaluate the system performance, we developed a testing approach that utilizes an airflow containing a precisely metered amount of pollen particles surrounded by a sheath flow to achieve the generation and lossless transmission of standard gas samples. We studied various factors affecting the pollen capture efficiency, including the applied voltage, air flow rate and humidity. Under optimized conditions, the system was successfully used in the measurement of airborne pollen particles within a wide range of concentrations, spanning 3 orders of magnitude.
更多
查看译文
关键词
Pollen measurement, Electrostatic enrichment, Machine vision, Automated analyzer
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要