Safety and Efficacy of Four Different Diagnostic Catheter Curves Dedicated to One-Catheter Technique of Transradial Coronaro-Angiography-Prospective, Randomized Pilot Study. TRACT 1: Trans RAdial CoronaryAngiography Trial 1

JOURNAL OF CLINICAL MEDICINE(2021)

引用 3|浏览18
暂无评分
摘要
Transradial coronaro-angiography (TRA) can be performed with one catheter. We investigate the efficacy of four different DxTerity catheter curves dedicated to the single-catheter technique and compare this method to the standard two-catheter approach. For this prospective, single-blinded, randomized pilot study, we enrolled 100 patients. In groups 1, 2, 3, and 4, the DxTerity catheters Trapease, Ultra, Transformer and Tracker Curve, respectively, were used. In group 5 (control), standard Judkins catheters were used. The study endpoints were the percentage of optimal stability, proper ostial artery engagement and a good quality angiogram, the duration of each procedure stage, the amount of contrast, and the radiation dose. The highest rate of optimal stability was observed in groups 2 (90%) and 5 (95%). Suboptimal results with at least one episode of catheter fallout from the ostium were most frequent in group 1 (45%). The necessity of using another catheter was observed most frequently in group 4. The analysis of time frames directly depending on the catheter type revealed that the shortest time for catheter introduction and for searching coronary ostia was achieved in group 2 (Ultra). There were no differences in contrast volume and radiation dose between groups. DxTerity catheters are suitable tools to perform TRA coronary angiography. The Ultra Curve catheter demonstrated an advantage over other catheters in terms of its ostial stability rate and procedural time.

更多
查看译文
关键词
transradial coronaro-angiography, single-catheter technique, coronary artery disease
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要