Fast And Accurate Polarimetric Calibration Of Infrared Imaging Polarimetric Sensors

APPLIED OPTICS(2021)

引用 3|浏览1
暂无评分
摘要
Polarimetric imaging cameras require polarimetric calibration to accurately estimate the incident Stokes vector of incoming radiation. This calibration establishes a relationship between changes in the sensor signal and incident Stokes vector. In the standard procedure, an imager is presented with a set of input Stokes vectors with two different radiance values. In the long-wavelength infrared (LWIR) and mid-wavelength infrared bands, blackbodies with different temperatures are used for each set of Stokes vectors. The radiometric offset is subtracted, and standard radiometric or nonuniformity correction procedures are performed in a separate step. This paper proposes an alternative all-in-one approach that combines radiometric calibration, nonuniformity correction, and polarimetric calibration. The standard and proposed methods are compared for a division-of-time LWIR polarimeter. The proposed calibration method achieves an RMS error of 0.34% compared with the conventional technique's error of 0.83%, yielding a factor of 2.4 improvement in the reconstructed accuracy of a linear Stokes vector; in addition, it is less time-consuming and less prone to ambient temperature fluctuations than the typical two-point method. The method also accounts for beam wander and narcissus effects and enables simple, straightforward polarimetric measurement. (C) 2021 Optical Society of America
更多
查看译文
关键词
accurate polarimetric calibration,infrared imaging,sensors