Developing an ultra-performance liquid chromatography-tandem mass spectrometry for detecting aldosterone in human plasma
JOURNAL OF CLINICAL LABORATORY ANALYSIS(2021)
摘要
Background Accurately measuring plasma aldosterone concentration is difficult but meaningful for primary aldosteronism (PA) diagnosis. Methods In this study, we developed an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for plasma aldosterone detection, evaluated its performance according to guidelines issued by CLSI, including detection limit, linearity, precision, and compared it with chemiluminescence immunoassay. Then, a reference range of plasma aldosterone in young people was established by using this method. Results The lower limit of quantitation (LOQ) was 10 pg/ml. The mean recovery rates of analyte added to serum were 100.07-102.05% in different concentrations. The linearity range was 20-2000 pg/ml. Inter-assay CVs were 2.20-3.97% at aldosterone concentrations of 65.66-854.75 pg/ml. The regression equation of UPLC-MS/MS (x) and chemiluminescence immunoassay (y) was y = 1.002x + 65.854 (r = 0.9456, n = 237). The reference range of plasma aldosterone detected by UPLC-MS/MS was 11.30-363.82 pg/ml in young people in South China, and there was no statistically significant difference in plasma aldosterone concentration between two genders. Conclusion In conclusion, UPLC-MS/MS can rapidly and accurately detect plasma aldosterone and is appropriate for clinical application.
更多查看译文
关键词
aldosterone, performance, plasma, primary aldosteronism, UPLC-MS, MS