A study on the sensitivity of the measurement of liquid density at different scattering angles using a gamma scattering technique.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine(2021)

引用 1|浏览3
暂无评分
摘要
This study aims to evaluate the sensitivity of the measurement of liquid density at different scattering angles using a gamma scattering technique. To perform this, the linear calibration curves of the ratio R (R is the ratio of the area under a single scattering peak for a liquid relative to that for water) versus the liquid density were constructed at different scattering angles. The sensitivity of the measurement is defined as the slope coefficient of these linear calibration curves. The results obtained from the Monte Carlo simulation data showed that the sensitivity of the measurement at different scattering angles including 70°, 80°, 90°, 100°, 110°, 120°, 130°, and 140° changes in the range from 0.44 to 0.48. Also, the results obtained from the experiment when performing the measurements at scattering angles of 90°, 100°, 110°, and 120° ranged from 0.46 to 0.48. This confirms that the dependence of the sensitivity of the measurement on scattering angle is insignificant. Besides, for every experimental dataset, we used each of 8 above-obtained calibration curves, in turn, to determine the densities of 8 liquids which yield the relative deviation between the measured density and the reference one is mostly less than 5%, the relative deviation of remaining cases (64 of 256 measurements) is in the range of 5.0%-9.9%.