Precision And Accuracy Of Fev1 Measurements From The Vitalograph Copd-6 Mini-Spirometer In A Healthy Ugandan Population

PLOS ONE(2021)

Cited 3|Views2
No score
Abstract
Objective Evaluate the accuracy and precision of the copd-6 mini-spirometer for FEV1 in a rural Ugandan population. Methods In a cross-sectional study, 171 smallholder farmers performed spirometry with copd-6, and a diagnostic-quality spirometer. Results and discussion The copd-6 underestimated FEV1 at low flows and overestimated FEV1 at high flows. Across all participants, the device slightly overestimated FEV1 by 0.04 [0.02; 0.06] L. Calibration data showed similar patterns. Conclusion The copd-6 could be considered as an affordable tool for research on lung function impairment in resource-constrained settings. However, further validation in a study population with obstructive lung disease is needed.
More
Translated text
Key words
fev1 measurements,mini-spirometer