Comparison of keratometry data using handheld and table-mounted instruments in healthy adults

INTERNATIONAL OPHTHALMOLOGY(2021)

引用 0|浏览1
暂无评分
摘要
Purpose To compare keratometry data between the handheld Retinomax K-plus 3 and the table-mounted IOLMaster 700. Methods Healthy adult volunteers were prospectively recruited to the study. All participants underwent 3 consecutive keratometry measurements using the Retinomax K-plus 3 and a single biometry assessment using the IOLMaster 700. Differences between the Retinomax K-plus 3 and the IOLMaster 700 were assessed using Wilcoxon test for paired samples, Spearman correlation, Bland–Altman and mountain plots. Result Twenty-eight healthy subjects with a median age of 37 years (interquartile range (IQR) 28–44 years) were included in the study. The median mean keratometry (mean K ) reading was higher using the Retinomax K-plus 3 (44.04D; IQR 42.96–45.61D) compared to the IOLMaster 700 (43.78D; IQR 43.22–44.90D, p < 0.01), with a mean difference of 0.18D (95% confidence interval (CI) 0.11–0.23D). Mean K readings were highly correlated between the 2 devices ( r = 0.995, p < 0.01). Bland–Altman plots showed 95% limits of agreement between −0.14D and 0.49D. Frequency histogram of mean K reading differences between the Retinomax K-plus 3 and the IOLMaster 700 showed that 56% of cases were between ± 0.2D, 93% of cases were between ± 0.4D and all cases were between ± 0.5D. Mean corneal astigmatism measurement was higher using the Retinomax K-plus 3 (1.01 ± 0.40D) compared to the IOLMaster 700 (0.77 ± 0.36D), with a mean difference of 0.23 ± 0.37D ( p < 0.01) between the devices. Conclusions A good agreement exists between the Retinomax K-plus 3 and the IOLMaster 700 regarding keratometry readings. This enables cataract surgeons to safely use the Retinomax K-plus 3 device when indicated.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要