Solid-State Neutron Detection Based On Methylammonium Lead Bromide Perovskite Single Crystals

ACS APPLIED MATERIALS & INTERFACES(2021)

引用 14|浏览7
暂无评分
摘要
Perovskite-based semiconductors, such as methylammonium and cesium lead halides (MPbX3: M = CH3NH3+ or Cs+; X = I-, Br-, or Cl-), have attracted immense attention for several applications, including radiation detection, due to their excellent electronic and optical properties.(1,2,3,4,5,6) In addition, the combination of perovskites with other materials enables unique device structures. For example, robust and reliable diodes result when combined with metal oxide semiconductors. This device can be used for detection of nonionizing and ionizing radiation. In this paper, we report a unique perovskite single-crystal-based neutron detector using a heterojunction diode based on single-crystal MAPbBr(3) and gallium oxide (Ga2O3) thin film. The MAPbBr(3)/Ga2O3 diodes demonstrate a leakage current of similar to 7 x 10(-10) A/mm(2), an on/off ratio of similar to 10(2), an ideality factor of 1.41, and minimal hysteresis that enables alpha particle, gamma-ray, and neutron detection at a bias as low as (-5 V). Gamma discrimination is further improved by 85% by optimizing the thickness of the perovskite single crystal. The MAPbBr(3)/Ga2O3 diodes also demonstrate a neutron detection efficiency of similar to 3.92% when combined with a B-10 neutron conversion layer.
更多
查看译文
关键词
perovskite single crystal, methylammonium lead bromide, CH3NH3PbBr3, MAPbBr(3), neutron detection, alpha detection, gamma detection, radiation detection