Biomechanical evaluation of intraosseous distal radioulnar joint prosthesis: A prosthesis designed based on Sauvé-Kapandji procedure

Orthopaedics & Traumatology: Surgery & Research(2021)

引用 4|浏览0
暂无评分
摘要
A B S T R A C T Introduction: To avoid the DRUJ surgical procedures disadvantages, a new intraosseous distal radioulnar prosthesis designed on Sauve-Kapandji procedure has been introduced. Stability of the prosthesis and biomechanics are to be evaluated in this article. Materials and methods: On a cadaveric study, during placement of the prosthesis, biometry of the bones, prosthesis stability (in axial and lateral tractions, wrist pronation and supination, and squeeze test), wrist range of motion before and after implantation, and radiographic evaluation were done on 16 cadavers. Results: Range of motion of the wrist joint before and after the insertion of the prosthesis, had no significant difference in all six directions. Stability of the prosthesis, when rotational pronation force was exerted, was greater than when rotational supination force was exerted. The prosthesis showed significant stability against longitudinal traction forces in a way that no prosthesis dislocation was observed up to 150 N forces. Stability of the prosthesis was investigated when lateral force was applied to different wrist positions. The most stable position of the prosthesis was in the case of lateral traction forces in supination where no case of dislocation was observed. Conclusions: The intraosseous distal radioulnar prosthesis demonstrated stable structure with no effect on wrist range of motion. Level of evidence: IV. (c) 2021 Elsevier Masson SAS. All rights reserved.
更多
查看译文
关键词
Distal radioulnar prosthesis,Cadaver study,Kapandji procedure
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要