[Influence of Reconstruction FOV and Matrix Size on the Quantitative Accuracy of FDG-PET: Comparison between OSEM and Bayesian Penalized Likelihood].

Shinya Mikami, Fumitoshi Nakamura,Akihiro Nakamura,Tatsuo Torizuka

Nihon Hoshasen Gijutsu Gakkai zasshi(2021)

Cited 0|Views3
No score
Abstract
The RC of OSEM images was not affected by changes in FOV, whereas the RC of BPL images decreased in small spheres, when FOV was 20 and 30 cm. The SUV of the OSEM images was not affected by the difference in pixel size. However, the SUV of BPL images degraded in the 1-mm pixel size; this influence was observed only when the FOV was changed. Conclusion: BPL images reconstructed using a small FOV might degrade the quantitative accuracy of small spheres.
More
Translated text
Key words
matrix size,positron emission tomography (PET),quantitative accuracy,reconstruction field of view