In vitro modeling accurately predicts cardiac lead fracture at 10 years

Heart Rhythm(2021)

引用 6|浏览3
暂无评分
摘要
BACKGROUND Development of a cardiac lead fracture model has the potential to differentiate well-performing lead designs from poor performing ones and could aid in future lead development.OBJECTIVE The purpose of this study was to demonstrate a predictive model for lead fracture and validate the results generated by the model by comparing them to observed 10-year implantable cardioverter-defibrillator lead fracture-free survival.METHODS The model presented here uses a combination of in vivo patient data, in vitro conductor fatigue test data, and statistical simulation to predict the fracture-free survival of cardiac leads. The model was validated by comparing the results to human clinical performance data from the Medtronic Sprint Fidelis (Minneapolis, MN) models 6931 (single coil, active fixation) and 6949 (dual coil, active fixation), as well as the Quattro model 6947 (dual coil, active fixation).RESULTS Median patient age in the single coil Fidelis 6931 population (64 years) was less than in the dual coil Fidelis 6949 and Quattro populations (68 years). Modeled and observed fracturefree survival for Quattro (>97%) was superior to that for Fidelis (<94%). The modeled survival agreed with the observed fracturefree survival data. The average model error was 0.3% (SD 1.2%).CONCLUSION This model for cardiac lead fracture-free survival using in vivo lead bending measurements and in vitro bench testing can be used to predict lead performance as observed by alignment with field survival data.
更多
查看译文
关键词
Cardiac lead fracture,Cardiac lead modeling,Cardiac lead survival,Fatigue strength,Implantable cardioverter-defibrillator leads
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要