Characterization of a novel large area microdosimeter system for low dose rate radiation environments

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2021)

引用 3|浏览29
暂无评分
摘要
A feasibility study is presented on a newly developed microdosimetry system named Octobox, for its application in low dose rate, mixed radiation environments. A full characterization of the device was performed at the Heavy Ion Medical Accelerator in Chiba (HIMAC), Japan, out-of-field of various heavy ion radiation fields — 290 MeV/u 12C, 230 MeV/u and 490 MeV/u 28Si and 400 MeV/u 20Ne ions, as well as a low dose rate 222Rn environment at the Australian Nuclear Sciences and Technology Organization (ANSTO). The device was shown to collect adequate statistics in a short period of time when compared to the MicroPlus probe with a single microdosimeter, while accurately measuring microdosimetric quantities and the corresponding average quality factor (Q¯) and dose equivalent (H) of the mixed radiation field. Good agreement of the microdosimetric spectra was also shown with Geant4 simulations for all presented ion fields. Based on the findings in this study, the Octobox is capable of being applied in mixed, low dose rate, radiation environments such as those encountered in space and aviation, as well as in underground mines for radiation protection purposes.
更多
查看译文
关键词
Large area SOI detector,Low dose rate,Microdosimetry,Silicon on insulator,Radiation protection