Surrogate Modeling-Based Acceleration of Multi-Harmonic Near-Field Measurements

2021 96th ARFTG Microwave Measurement Conference (ARFTG)(2021)

Cited 1|Views2
No score
Abstract
In this paper, a surrogate modeling-based acceleration technique for multi-harmonic phase-coherent electrooptic near-field measurements is presented. The implementation uses an adaptive sampling and modeling algorithm instead of the conventional raster scanning approach, which reduces the measurement time by a factor of 9, from 7 hours to 45 minutes, and the number of samples by a factor of 23, fr...
More
Translated text
Key words
Microwave measurement,Accelerometers,Adaptation models,Measurement errors,Phase measurement,Fluctuations,Measurement uncertainty