Continuous Glucose Monitoring In The Intensive Care Unit Following Total Pancreatectomy With Islet Autotransplantation In Children: Establishing Accuracy Of The Dexcom G6 Model
JOURNAL OF CLINICAL MEDICINE(2021)
摘要
Hyperglycemia is detrimental to postoperative islet cell survival in patients undergoing total pancreatectomy with islet autotransplantation (TPIAT). This makes continuous glucose monitoring (CGM) a useful management tool. We evaluated the accuracy of the Dexcom G6 CGM in pediatric intensive care unit patients following TPIAT. Twenty-five patients who underwent TPIAT had Dexcom G6 glucose values compared to paired serum glucose values. All paired glucose samples were obtained within 5 minutes of each other during the first seven days post TPIAT. Data were evaluated using mean absolute difference (MAD), mean absolute relative difference (MARD), %20/20, %15/15 accuracy, and Clarke Error Grid analysis. Exclusions included analysis during the CGM "warm-up" period and hydroxyurea administration (known drug interference). A total of 183 time-matched samples were reviewed during postoperative days 2-7. MAD was 14.7 mg/dL and MARD was 13.4%, with values of 15.2%, 14.0%, 12.1%, 11.4%, 13.2% and 14.1% at days 2, 3, 4, 5, 6 and 7, respectively. Dexcom G6 had a %20/20 accuracy of 78%, and a %15/15 accuracy of 64%. Clarke Error Grid analysis showed that 77% of time-matched values were clinically accurate, and 100% were clinically acceptable. The Dexcom G6 CGM may be an accurate tool producing clinically acceptable values to make reliable clinical decisions in the immediate post-TPIAT period.
更多查看译文
关键词
continuous glucose monitoring, total pancreatectomy with islet autotransplantation, mean absolute relative difference, mean absolute difference, Dexcom G6
AI 理解论文
溯源树
样例
![](https://originalfileserver.aminer.cn/sys/aminer/pubs/mrt_preview.jpeg)
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要