In-field and out-of-field microdosimetric characterisation of a 62 MeV proton beam at CATANA

MEDICAL PHYSICS(2021)

引用 4|浏览8
暂无评分
摘要
Purpose A 5 and 10 mu m thin silicon on insulator (SOI) 3D mushroom microdosimeter was used to characterize both the in-field and out-of-field of a 62 MeV proton beam. Methods The SOI mushroom microdosimeter consisted of an array of cylindrical sensitive volumes (SVs), developed by the Centre for Medical Radiation Physics, University of Wollongong, was irradiated with 62 MeV protons at the CATANA (Centro di AdroTerapia Applicazioni Nucleari Avanzate) facility in Catania, Italy, a facility dedicated to the radiation treatment of ocular melanomas. Dose mean lineal energy, (yD over bar ), values were obtained at various depths in PMMA along a pristine and spread out Bragg peak (SOBP). The measured microdosimetric spectra at each position were then used as inputs into the modified Microdosimetric Kinetic Model (MKM) to derive the RBE for absorbed dose in a middle of the SOBP 2Gy (RBED). Microdosimetric spectra were obtained with both the 5 and 10 mu m 3D SOI microdosimeters, with a focus on the distal part of the BP. The in-field and out-of-field measurement configurations along the Bragg curve were modeled in Geant4 for comparison with experimental results. Lateral out-of-field measurements were performed to study secondary particles' contribution to normal tissue's dose, up to 12 mm from the edge of the beam field, and quality factor and dose equivalent results were obtained. Results Comparison between experimental and simulation results showed good agreement between one another for both the pristine and SOBP beams in terms of yD over bar and RBED. Though a small discrepancy between experiment and simulation was seen at the entrance of the Bragg curve, where experimental results were slightly lower than Geant4. The dose equivalent value measured 12 mm from the edge of the target volume was 1.27 +/- 0.15 mSv/Gy with a Q over bar value of 2.52 +/- 0.30, both of which agree within uncertainty with Geant4 simulation. Conclusions These results demonstrate that SOI microdosimeters are an effective tool to predict RBED in-field as well as dose equivalent monitoring out-of-field to provide insight to probability of second cancer generation.
更多
查看译文
关键词
microdosimetry, out-of-field, RBE, Silicon-On-Insulator (SOI)
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要