Efficacy And Safety Of Satake Hotballoon (R) Catheter For Treatment Of Paroxysmal Atrial Fibrillation-A Post-Marketing Surveillance Study

CIRCULATION JOURNAL(2021)

引用 5|浏览9
暂无评分
摘要
Background: SATAKE HotBalloon (R) catheter (HBC) is a radiofrequency balloon catheter for the treatment of atrial fibrillation (AF), and was approved for use in Japan to treat drug-resistant paroxysmal AF in 2015. Post-marketing surveillance study was conducted by Toray Industries, Inc. to evaluate the efficacy and safety of HBC treatment in patients with paroxysmal AF in a real-world setting. This study is the first nation-wide survey of HBC treatment for paroxysmal AF in clinical practice in Japan.Methods and Results: This was a single-arm, multicenter observational study with an observation period of 48 weeks after ablation. Pulmonary vein isolation and AF non-recurrence rates were evaluated and adverse events (AEs) were observed at 46 sites in Japan. An AF event was defined as recurrence of AF or re-ablation from 12 to 48 weeks after ablation. The success rate of pulmonary vein isolation was 99.0% (486/491) for patients with AF. The cumulative AF non-recurrence rate was 94.1% at 24 weeks and 87.8% at 48 weeks. AEs were found to occur 21.5% (114/530), and ablation-related AEs were found to occur 2.6% (14/530) during the study period, with the most common being pericardial effusion (0.8%, 4/530).Conclusions: This study demonstrates the efficacy and safety of HBC ablation in Japanese patients with recurrent symptomatic paroxysmal AF refractory to antiarrhythmic therapy.
更多
查看译文
关键词
Atrial fibrillation, Catheter ablation, HotBalloon (R), Pulmonary vein isolation