Absolute Response Of A Fuji Bas-Tr Imaging Plate To Low-Energy Protons (< 0.2 Mev) And Carbon Ions (< 1 Mev)

REVIEW OF SCIENTIFIC INSTRUMENTS(2021)

引用 1|浏览34
暂无评分
摘要
This paper reports on the absolute response of a Fuji BAS-TR image plate to relatively low-energy protons (<0.2 MeV) and carbon ions (<1 MeV) accelerated by a 10-TW-class compact high-intensity laser system. A Thomson parabola spectrometer was used to discriminate between different ion species while dispersing the ions according to their kinetic energy. Ion parabolic traces were recorded using an image plate detector overlaid with a slotted CR-39 solid-state detector. The obtained response function for the protons was reasonably extrapolated from previously reported higher-ion-energy response functions. Conversely, the obtained response function for carbon ions was one order of magnitude higher than the value extrapolated from previously reported higher-ion-energy response functions. In a previous study, it was determined that if the stopping range of carbon ions is comparable to or smaller than the grain size of the phosphor, then some ions will provide all their energy to the binder resin rather than the phosphor. As a result, it is believed that the imaging plate response will be reduced. Our results show good agreement with the empirical formula of Lelasseux et al., which does not consider photo-stimulated luminescence (PSL) reduction due to the urethane resin. It was shown that the PSL reduction due to the deactivation of the urethane resin is smaller than that previously predicted.
更多