Plan quality effects of maximum monitor unit constraints in pencil beam scanning proton therapy for central nervous system and skull base tumors.

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology(2021)

引用 0|浏览18
暂无评分
摘要
PURPOSE/OBJECTIVE(S):With reports of CNS toxicity in patients treated with proton therapy at doses lower than would be expected based on photon data, it has been proposed that heavy monitor unit (MU) weighting of pencil beam scanning (PBS) proton therapy spots may potentially increase the risk of toxicity. We evaluated the impact of maximum MU weighting per spot (maxMU/spot) restrictions on PBS plan quality, prior to implementing clinic-wide maxMU/spot restrictions. MATERIALS/METHODS:PBS plans of 11 patients, of which 3 plans included boosts, for a total of 14 PBS sample cases were included. Per sample case, a single dosimetrist created 4 test plans, gradually reducing the maxMU/spot in the plan. Test Plan 1, unrestricted in maxMU/spot, was the reference for all restricted plan comparisons (comparison sets 2 vs. 1; 3 vs. 1; and 4 vs. 1). The impact of MU/spot restrictions on plan quality metrics were analyzed with Wilcoxon signed rank test analyses. Treatment delivery time was modeled for a representative case. RESULTS:A total of 14 PBS sample cases, 7 (50%) single-field optimized, 7 (50%) multi-field optimized, 9 (64%) delivering > 3500 cGy, 9 (64%) with 3 beams, and 7 (50%) without a range shifter were included. There were no differences in plan quality metrics of target coverage (V95% and V100% prescription), conformality and gradient indices, hot spot volume (V105% prescription), and dose to normal brain (V10%/30%/50%/70%/90%/100% prescription) with reductions of allowable maxMU/spot across all comparison sets (p > 0.05). Max MU/spot restrictions did not increase treatment delivery time when analyzed for a representative case. CONCLUSION:MaxMU/spot restrictions within the thresholds evaluated in this study did not degrade overall plan quality metrics. Future studies should evaluate spot weighting with linear energy transfer/relative biologic effectiveness-informed planning to determine if spot weighting manipulation impacts clinical outcomes and mitigates toxicity.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要