An accurate analytical method for the determination of cadmium: Ultraviolet based photochemical vapor generation-slotted quartz tube based atom trap-flame atomic absorption spectrophotometry

Measurement(2021)

引用 8|浏览2
暂无评分
摘要
•PVG method was firstly coupled with T-SQT-AT method prior to FAAS determination.•LOD and LOQ values for the proposed method were 2.5 and 8.2 µg/L, respectively.•Sensitivity of the FAAS system was enhanced about 106 folds.•Recovery results were ranged from 87.6 to 105.3% for the samples.
更多
查看译文
关键词
Cadmium,Atomic absorption spectrometry,Photochemical vapor generation,T-shaped slotted quartz tube,Atom trapping