Impact of CT-based planning and bladder volume changes on the dosimetry of vaginal vault brachytherapy

Journal of Radiotherapy in Practice(2021)

引用 1|浏览2
暂无评分
摘要
Aim: To compare dose to target and organs at risk (OARs) in conventional plan (2D) versus computed tomography (CT)-based three-dimensional (3D) plan in vaginal cuff brachytherapy (VBT) and to compare the effect of bladder distension on target and OARS dosimetry. Materials and methods: Post-hysterectomy patients with an indication for VBT were included in the study. All patients underwent planning CT scans with a full bladder and an empty bladder protocol. For each CT, two plans were generated-one library-based 2D plan and another CT-based 3D plan. Dosimetric parameters were recorded for clinical target volume (CTV) and OARs. Results: A total of 92 observations were made from data collected from 46 patients. Difference between CTV dose in terms of 2D and 3D plans were not statistically significant for CTV (p = 0.11). Significant reduction in D0.1cc, D1cc and D2cc dose parameters were observed in bladder, rectum, sigmoid and bowel doses with the 3D plan (p < 0.001). Bladder distension showed a 20% reduction in dose for bowel (p < 0.001). Bladder distension also showed a 6.12% (p = 0.047) increase in D2cc, but there was a significant reduction in the mean dose to the bladder. Conclusion: Our study demonstrates the dosimetric benefits with 3D CT-based planning for VBT over 2D-based conventional planning and benefit of bladder distension in the reduction of bowel dose without compromising dose to the target volume.
更多
查看译文
关键词
bladder, CT-based planning, gynaecological cancer, Vaginal brachytherapy