Impending Trunnion Failure: An Uncommon Radiographic Presentation of Total Hip Arthroplasty Failure

Emma Baker,Cristy French, Pamela Brian, Jonelle Thomas,Charles M. Davis III

ARTHROPLASTY TODAY(2021)

引用 1|浏览3
暂无评分
摘要
Trunnionosis is emerging as an early mode of failure in conventional metal-on-polyethylene total hip arthroplasty. It is defined as wear or corrosion at the trunnion, the taper at the femoral head-neck interface. Trunnion wear can result in a variety of negative sequelae and, in severe cases, necessitate revision arthroplasty. We describe a 64-year-old man with a metal-on-polyethylene total hip arthroplasty who presented with a sensation of clunking in the hip. Initial imaging and laboratory studies were inconclusive, and the decision was made to monitor. Two years later, trunnion wear was detected on radiographs, presenting as an abnormal alignment of the femoral neck relative to the femoral head. Several case reports and series describe catastrophic total hip arthroplasty failure due to trunnionosis. However, few describe the radiographic signs of wear at the trunnion before gross failure. This early presentation is important to recognize to minimize patient morbidity and aid surgical planning. (c) 2020 The Authors. Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/ licenses/by-nc-nd/4.0/).
更多
查看译文
关键词
Trunnionosis,Trunnion failure,Trunnion corrosion,Total hip arthroplasty,Adverse local tissue reaction