Subcutaneous Implantable Cardioverter-Defibrillator Electrode Fracture: Follow-Up, Troubleshooting, And Evaluation

Journal of cardiovascular electrophysiology(2021)

引用 2|浏览0
暂无评分
摘要
The subcutaneous-implantable cardioverter-defibrillator (S-ICD) and its electrode were developed to avoid long-term complications of transvenous leads in the vasculature. We report a case of unexpected, inappropriate S-ICD shocks due to oversensing of high-amplitude, nonphysiologic, electrical noise artifacts that were not preceded by high-impedance alerts or sensing electrogram noise detections. Following explant, high-magnification X-ray imaging of the S-ICD electrode demonstrated partial fracture of the distal sensing conductor located near a short radius bend in the electrode at the electrode-header interface. Clinicians should be aware of a potential for fatigue failure fracture of the S-ICD electrode. Recommendations for systematic S-ICD follow-up and troubleshooting are discussed.
更多
查看译文
关键词
electrode fracture, inappropriate ICD therapy, oversensing, subcutaneous ICD