A Stop-Timing Microchannel Plate Detector For Multi-Reflection Time-Of-Flight Mass Analyzer

JOURNAL OF INSTRUMENTATION(2021)

引用 0|浏览19
暂无评分
摘要
Astop-timing microchannel detector with an integrated fast amplifier has been developed and its mechanical and electronic structure described in detail. A typical output amplitude of 1.1V, rise time of 10 ns, noise amplitude of 2.5mV(rms) and timing jitter of 23 ps of the detector have been obtained. The detector has tested with Cs-133(1+) ions in the multi-reflection time-of-flight mass analyzer at the Spectrometer for Heavy Atom and Nuclear Structure (SHANS) and a mass resolving power of 19500 has been achieved for a total time-of-flight of 5.85 ms.
更多
查看译文
关键词
Heavy-ion detectors, Timing detectors, Very low-energy charged particle detectors