谷歌Chrome浏览器插件
订阅小程序
在清言上使用

Surface Dose Accuracy In Vmat Head And Neck Radiation Treatment Using Bolus

10TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE)(2019)

引用 1|浏览10
暂无评分
摘要
The accuracy of Eclipse (TM)-calculated head and neck VMAT plan dose delivery was assessed in the situation when there is a gap between placed bolus on a head and neck thermoplastic mask and the body surface. GafchromicEBT3 film dosimetry was used for dose delivery validation. The results from 3%, 3 mm 2D gamma comparison indicate better than 95% agreement between measured film dose and calculated plan dose for gap widths up to 25 mm between placed bolus and body surface.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要