Reciprocating-Flowing On-A-Chip Enables Ultra-Fast Immunobinding For Multiplexed Rapid Elisa Detection Of Sars-Cov-2 Antibody

BIOSENSORS & BIOELECTRONICS(2021)

Cited 27|Views12
No score
Abstract
The worldwide epidemic of novel coronavirus disease (COVID-19) has led to a strong demand for highly efficient immunobinding to achieve rapid and accurate on-site detection of SARS-CoV-2 antibodies. However, hour-scale time-consumption is usually required to ensure the adequacy of immunobinding on expensive large instruments in hospitals, and the common false negative or positive results often occur in rapid on-site immunoassay (e.g. immunochromatography). We solved this dilemma by presenting a reciprocating-flowing immunobinding (RFimmunobinding) strategy. RF-immunobinding enabled the antibodies in fluid contacting with the corresponding immobilized antigens on substrate repeatedly during continuous reciprocating-flowing, to achieve adequate immunobinding within 60 s. This strategy was further developed into an immunoassay method for the serological detection of 13 suspected COVID-19 patients. We obtained a 100% true negative and true positive rate and a limit of quantification (LOQ) of 4.14 pg/mL. Our strategy also can be a potential support for other areas related to immunorecognition, such as proteomics, immunopharmacology and immunohistochemistry.
More
Translated text
Key words
Coronavirus disease, Microfluidic chip, Immunoassay, Interaction dynamics, Multiplexed analysis
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined