Leak-free integrated microfluidic channel fabrication for surface plasmon resonance applications

JOURNAL OF MICROMECHANICS AND MICROENGINEERING(2020)

Cited 4|Views14
No score
Abstract
In this paper, we describe a novel fabrication method of a microfluidic integrated surface plasmon resonance (SPR) gold chip based on a (3-mercaptopropyl) trimethoxy silane (MPTMS) self-assembled monolayer. This monolayer was formed at the surface of a microfluidic chip made of polydimethylsiloxane (PDMS). Its presence was confirmed by contact angle and Fourier transform infrared spectroscopy measurements on the modified PDMS surface. A basic, but nevertheless appropriate, 4-channel microfluidic system was made on PDMS and reported on a gold SPR sensor. Sealing tests were carried-out by injecting continuous flows of solutions under gradient pressure up to 1.8 bar. Bonding strength of chemical and corona binding were measured and compared. The test of the integrated microfluidic SPR sensor on an SPR bench validated its functionality and proved as well that no leakage is observed between the different microfluidic channels.
More
Translated text
Key words
PDMS-gold bonding,surface functionalization,MPTMS,silanes,surface plasmon resonance,microfluidic-chip,tensile strength