A Novel Platform Of Using Copper (Ii) Complex With Triazole-Carboxilated Modified As Bidentated Ligand Spce For The Detection Of Hydrogen Peroxide In Milk
JOURNAL OF ELECTROANALYTICAL CHEMISTRY(2020)
Abstract
A screen-printed carbon electrode (SPCE) modified with copper (II) 1-(3-bromobenzyl)-5-methyl-1H-1,2,3-triazole-4-carboxylate [Cu(bmtc)(2)(H2O)] and was used as an electrochemical sensor of hydrogen peroxide. The material was characterised by crystal X-ray and cyclic voltammetry. The electrochemical behaviour of the modified electrode [Cu(bmtc)(2)(H2O)]/SPCE was evaluated against bare SPCE for the reduction of 1.0 mM H2O2 in phosphate buffer (pH = 7.0), where the modified electrode responded similar to -0.55 V even though the SPCE did not present response. On the other hand, analytical parameters such as reproducibility (2.7%), repeatability (2.6%), LOD (5.8 x 10(-7) mol L-1), selectivity (physiological substances) and stability were evaluated.
MoreTranslated text
Key words
[cu(bmtc)2(H2O)], Hydrogen peroxide, Milk, Amperometry, Electrochemical sensor