The Performance Of The Commissioned Notre Dame Multi-Reflection Time-Of-Flight Mass Spectrometer

B. Liu,M. Brodeur, D.P. Burdette, J.M. Kelly, T. Kim,J. Long,P.D. O’Malley

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2021)

引用 4|浏览12
暂无评分
摘要
The future N = 126 factory at Argonne National Laboratory is aimed to produce neutron-rich elements around the N = 126 peak of the rapid-neutron capture process using multi-nucleon transfer reactions. However, this radioactive ion beam production method will entail the delivery of contaminant ions. To remove the isobar contaminants from the beam, a multi-reflection time-of-flight mass spectrometer has been built and successfully commissioned at the University of Notre Dame, reaching a resolving power of 70,000, with an efficiency of 10%. According to stability study conducted, the resolving power is potentially limited to 83,000 by the power supply stability, however it can be mitigated by the addition of RC filtration as done for other MR-TOFs. The commissioned MR-TOF has been transported to Argonne National Laboratory to be used in the N = 126 factory.
更多
查看译文
关键词
Multi-reflection time-of-flight mass spectrometer, Isobar separator, Ion trap, Radioactive ion beams