Performance Characteristics Of A Rapid Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Detection Assay At A Public Plaza Testing Site In San Francisco

JOURNAL OF INFECTIOUS DISEASES(2021)

引用 118|浏览25
暂无评分
摘要
We evaluated the performance of the Abbott BinaxNOW rapid antigen test for coronavirus disease 2019 (Binax-CoV2) to detect virus among persons, regardless of symptoms, at a public plaza site of ongoing community transmission. Titration with cultured severe acute respiratory syndrome coronavirus 2 yielded a human observable threshold between 1.6 x 10(4)-4.3 x 10(4) viral RNA copies ( cycle threshold [Ct], 30.3-28.8). Among 878 subjects tested, 3% (26 of 878) were positive by reverse-transcription polymerase chain reaction, of whom 15 of 26 had a Ct <30, indicating high viral load; of these, 40% (6 of 15) were asymptomatic. Using this Ct threshold (<30) for Binax-CoV2 evaluation, the sensitivity of Binax-CoV2 was 93.3% (95% confidence interval, 68.1%-99.8%) (14 of 15) and the specificity was 99.9% (99.4%-99.9%) (855 of 856).
更多
查看译文
关键词
COVID-19, SARS-CoV-2, Rapid Antigen Test, Point of Care testing