B-line Detection using Amplitude Modulation-Frequency Modulation (AM-FM) Features

Proceedings of SPIE(2018)

引用 0|浏览7
暂无评分
摘要
Pneumonia is one of the most common acute respiratory infections among pediatric populations worldwide. Ultrasound is becoming increasingly important in the diagnosis of lung diseases as a more portable and safer alternative to X-ray imaging. In the current work, we present a new automatic system for detection of B-lines, one of the distinctive features of pneumonic ultrasound scans, using amplitude modulation-frequency modulation (AM-FM). Features are evaluated on 109 videos obtained from 100 pediatric patients using a Verasonics V1 scanner. Further, the results were compared to the ones obtained with a previously published spectral feature (SF) method. Sensitivities of 92% and 83% and specificities of 91% and 70% were obtained on zone-1 and zone-2 of the lungs, respectively. In contrast, the SF method provided sensitivities of 72% and 68% in zone-1 and zone-2, respectively, and specificities of 68% and 46% in zone-1 and zone-2, respectively. In addition, the AM-FM method allowed increasing the F1-score when compared to the SF method from 70% to 87% and from 61% to 78% in zone-1 and zone-2, respectively. The results suggest the proposed method may be useful for the computer assisted diagnosis of pneumonia.
更多
查看译文
关键词
pneumonia,AM-FM analysis,lung ultrasound
AI 理解论文