Differential Reference Electrode Applied to Arrayed pH Sensor and Arrayed Glucose Biosensor Based on Microfluidic Framework

Lecture Notes in Engineering and Computer Science(2013)

引用 0|浏览0
暂无评分
摘要
In this study, the radio frequency (R. F.) sputtering system and screen printing technology were used to fabricate the arrayed pH sensor. And the glucose oxidase (GOx) enzyme was immobilized on the arrayed pH sensor as the arrayed glucose biosensor. Furthermore, the computer numerical control (CNC) technique was used to fabricate the microfluidic mold. The differential reference electrode can provide a stable potential in different pH solutions, which was applied in arrayed pH sensor and arrayed glucose biosensor. The arrayed pH sensor and arrayed glucose biosensor were integrated to the poly-dimethylsiloxane (PDMS) microchannel as the microfluidic device for observing the characteristics of the arrayed pH sensor and arrayed glucose biosensor at dynamical conditions. At dynamical conditions with flow rate controlled at 20 ml/min, the average sensitivity and linearity of the arrayed pH sensor were 57.27 mV/pH and 0.986, respectively. And at the flow rate controlled at 5 ml/min, the average sensitivity and linearity of the arrayed glucose biosensor were 9.67 mV(100 mg/dL)(-1) and 0.943, respectively.
更多
查看译文
关键词
Arrayed pH sensor,glucose oxidase enzyme,immobilized,arrayed glucose biosensor,microfluidic mold,differential reference electrode