Guide sheath breakage during endobronchial ultrasonography

RESPIROLOGY CASE REPORTS(2021)

引用 0|浏览0
暂无评分
摘要
This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the peripheral bronchus near the tumour. The band could not be removed endoscopically. As the tumour was diagnosed as a colon cancer metastasis, we performed a partial lung resection to remove the RB and nodule together four months after bronchoscopy.
更多
查看译文
关键词
EBUS-GS,endobronchial ultrasonography,guide sheath breakage,instrumental breakage