Guide sheath breakage during endobronchial ultrasonography
RESPIROLOGY CASE REPORTS(2021)
摘要
This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the peripheral bronchus near the tumour. The band could not be removed endoscopically. As the tumour was diagnosed as a colon cancer metastasis, we performed a partial lung resection to remove the RB and nodule together four months after bronchoscopy.
更多查看译文
关键词
EBUS-GS,endobronchial ultrasonography,guide sheath breakage,instrumental breakage
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要