Energy Calibration of Plastic Scintillator Detector

AIP Conference Proceedings(2019)

引用 2|浏览2
暂无评分
摘要
This paper is devoted to a plastic scintillator detector calibration method. We calibrated the large volume polystyrene gamma detector SPD200.200.100 by NUVIA Group. The essential issue of the plastic scintillator is that Compton scattering is the dominant effect of photon interaction within the detector and it is no apparent full energy absorption peak in a spectrum. Our approach is based on MCNP code simulation where the detector was modeled. Then measured and calculated spectra are compared to determine the Gaussian energy broadening effect. Finally, the measured spectra can be calibrated using Compton maximum and Compton edge energy. The results of this study show that measured and calculated spectra are in good agreement.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper